36 5 System operation
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
5.3 System start-up
Switching ON the ventilator
Switch ON the ventilator
Use the start /stop button on
fabian +nCPAP evolution to power
ON the device.
When the device is powered ON,
the Start-up screen will appear
and the device will perform a self-
test.
The SOFTWARE REVISION, the
CHECK SUMS and the status of the
interfaces will be checked.
(ok = no error)
Loudspeaker (audio) check
After a successful power ON, the
device will prompt the user to
perform the acoustic audio alarm
test
In order to proceed, you must
acknowledge that the alarm is
audible, continue using the
ventilator without audio alarm or
shut down the ventilator.
The O
2
will automatically be calibrated
after leaving the Calibration menu and
repeated once every 24 hours.