42 5 System operation
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
5.7 Emergency shutdown
WARNING
In the event of a fault resulting in loss of normal control the emergency shutdown
procedure may be used. This procedure disconnects all power to the ventilator and any
other devices connected to the ventilator.
Before performing this procedure, ensure that the alternate ventilation method (for
example manual resuscitation) is present, ready and standing by.
Emergency shutdown
Remove patient from the
ventilator.
Press and hold the
start / stop button (1)
for one to two seconds
then release.
After releasing, press and hold the
start / stop button (1) again for
five seconds.
After the device has powered OFF, a
notice tone will sound.
Press the Alarm Silence button to confirm system switch OFF.
Hold the button for at least three seconds until the WARNING LED
stops.
1