70 8 Battery operation
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
8 Battery operation
The device features an Internal Battery.
Battery run time in IPPV mode on full charge is approximately 3 hours.
These values may vary depending on the parameter settings. The Battery charge is continuously
monitored. The device continuously calculates the remaining Run Time and displays it in the
Information bar.
WARNING: A reduction in Battery remaining run time when changing the
Operating mode can cause an Automatic shutdown of ventilator!
Changing the Operating mode can greatly shorten the battery run time.
Monitor the remaining Battery run time when changing the settings.
NEVER leave the device and patient unattended during battery
operation.
Establish mains power supply in good time!
In the event of Power Failure, the pneumatic system will automatically open
the safety valve to atmospheric pressure so that pressure can’t build in the
breathing system, allowing spontaneous patient breathing.
When storing the ventilator for long term storage, please remove the Battery
pack.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals