121 19 Technical specifications
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
19 Technical specifications
19.1 Ambient conditions
During operation:
Temperature
Air Pressure
Relative Humidity
10 to 40°C (50 to 104 °F)
70 to 106 kPa
10 to 90%, non-condensing
During transportation:
Temperature
Air Pressure
Relative Humidity
-20 to 60°C (-4 to 140 °F)
50 to 106 kPa
10 to 95%, non-condensing
During storage:
Temperature
Air Pressure
Relative Humidity
0 to 4C (+3F to 104 °F)
70 to 106 kPa
20 to 80%, non-condensing
IMPORTANT:
For moving the device between ambient and controlled temperatures, removing the
device from storage, and before use; acclimate the fabian device for 12 hours.
The Minimum and Maximum Ambient temperatures in the clinical area will not affect the
accuracy of the Flow, Pressure or Oxygen measurement.
19.2 Monitoring
Acoustic Pressure of Alarm tones. Alarm tone sequence IEC 60601-1-8:
Lowest possible setting:
Highest possible setting:
52 db(A)
68 db(A)
A weighted sound pressure level in 1m distance:
40dB
Expiratory Minute Volume
Alarm Lower threshold Value:
Alarm Upper threshold Value:
If Lower Alarm Limit not reached
If Upper Alarm Limit exceeded
Airway Pressure
Alarm Lower threshold Value:
Alarm Upper threshold Value:
maximum airway pressure:
If Lower Alarm Limit not reached
If Upper Alarm Limit exceeded
100 mbar
Inspiratory O
2
concentration
Alarm:
Set Value ±5 Vol% for >1 minute at 21 Vol%
below threshold Value 18 Vol%
Breath Frequency
Alarm:
If Alarm Limit exceeded
Volume Monitoring
Alarm Lower Threshold Value:
If the breath volume set was not supplied
Apnea Alarm