5 System operation 39
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
What
How
Target
Leakage
Test
Start ventilation mode: CPAP
Enter the following settings:
CPAP:
5
mbar
Pmanual:
80
mbar
Press and hold the
Manual Breath button
Pressure of 80 ±4 mbar
is achieved on the pressure
graph
Caution:
Ensure that Pmanual and
other pressures are reduced
to safe levels prior to
connecting to patient.
Function
Test
Start Ventilation mode: IPPV
Enter the following settings:
Iflow:
10
L/min
Eflow:
8
L/min
PEEP:
5
mbar
Pinsp:
20
mbar
Freq. (Rate):
30
1 / min
Itime:
1
second
Oxygen:
30
vol.%
Monitor ventilation
parameters to ensure the
values are within the
following ranges.
P insp.:
20
±2 mbar
PEEP:
5
±1 mbar
O
2
:
30
±2 Vol %
Alarms
Disconnect the Inspiration tube and block
the Inspiratory port on the ventilator.
Alarm: Tube Occlusion
Disconnect the Expiration tube and block
the opening of the Expiratory tube.
Alarm: Tube Occlusion
Crimp and hold Proximal Measurement
tube.
Alarm: Tube Occlusion
Humidifier
Fill level
Function
Water fill level sufficient.
Function OK
CAUTION:
Ensure that Pmanual and other pressures are reduced to safe levels prior to connecting to
patient.