108 15 Sterilization / cleaning / disinfection
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
NOTE: Containers are recommended to transport used accessories between the NICU
and the cleaning/sterilization area.
2. Remove the exhalation membrane from the exhalation membrane holder:
2.1. Before cleaning the membrane holder and the membrane, disassemble the
exhalation membrane from the membrane holder by pulling out from the
second circumference of the exhalation valve membrane.
15.2.3.2 Preparation for sterilization by automated cleaning
1. Rinse under running tap cold water (<35°C / 95°F) to remove excess soil.
2. Prepare a detergent bath using Neodisher Mediclean Forte solution at the
manufacturer’s recommendation of 10 mL per Liter using utility (tap) water 20°C /
68°F - 30°C / 86°F.
3. Transfer the cassette items on to the 4-Level manifold rack accessory (or other
appropriate rack system) contained inside the washer for processing.
4. Select the appropriate cycle as listed below: