108 15 Sterilization / cleaning / disinfection
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
NOTE: Containers are recommended to transport used accessories between the NICU
and the cleaning/sterilization area.
2. Remove the exhalation membrane from the exhalation membrane holder:
2.1. Before cleaning the membrane holder and the membrane, disassemble the
exhalation membrane from the membrane holder by pulling out from the
second circumference of the exhalation valve membrane.
15.2.3.2 Preparation for sterilization by automated cleaning
1. Rinse under running tap cold water (<35°C / 9F) to remove excess soil.
2. Prepare a detergent bath using Neodisher Mediclean Forte solution at the
manufacturers recommendation of 10 mL per Liter using utility (tap) water 2C /
68°F - 3C / 86°F.
3. Transfer the cassette items on to the 4-Level manifold rack accessory (or other
appropriate rack system) contained inside the washer for processing.
4. Select the appropriate cycle as listed below:
Phase
Recirculation
Time (Minutes)
Temperature
Detergent Type and Concentration
(If applicable)
Pre-wash 1
02:00
Cold tap water
N/A
Wash 1
02:00
45°C / 11F
Tap water
(Set Point)
Neodisher Mediclean Forte 10 ml
per Liter
Neutralization
Wash
02:00
Warm tap water
Neodisher Z 1.0 ml per Liter
Rinse 1
01:00
Warm tap water
N/A
Dry Time
07:00
90°C / 19F
N/A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals