112 16 Setting ranges and parameters
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Mode
Parameter
SIMV
SIMV+PSV
Neonatal
Pediatric
Neonatal
Pediatric
min
max
min
max
min
max
min
max
Eflow [L/min]
1
32
1
32
1
32
1
32
Etime [sec]
0.5
30
0.5
30
0.5
30
0.5
30
Flush Time [sec]
0
120
0
120
0
120
0
120
Frequency (Rate) [1 / min]
2
100
2
100
2
100
2
100
Iflow [L/min]
1
32
1
32
1
32
1
32
Itime [sec]
0.1
2
0.3
2
0.1
2
0.3
2
Man. Breath Time [sec]
2
30
2
30
2
30
2
30
O
2
[%]
21
100
21
100
21
100
21
100
O
2
Flush [%]
23
100
23
100
23
100
23
100
PEEP [mbar]
0
30
0
30
0
30
0
30
Pinsp (backup) [mbar]
4
80
4
80
4
80
4
80
Pmax [mbar]
4
80
4
80
4
80
4
80
Ppsv [mbar]
2
80
2
80
PSV Termination criteria [%]
1
85
1
85
Rise Time [sec]
0.1
2
0.3
2
0.1
2
0.3
2
Trigger [Volume]
1
10
1
10
1
10
1
10
V guarant [mL]
0.8
60
10
300
0.8
60
10
300
V limit [mL]
1
150
10
500
1
150
10
500