114 16 Setting ranges and parameters
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Mode
Parameters
NCPAP
DUOPAP
O
2
Therapy
Neonatal
Neonatal
Neonatal
Pediatric
min
max
min
max
min
max
min
max
CPAP [mbar]
2
13
2
13
E-time [sec]
0.2
30
Flowmin [L/min]
0
15
0
30
Flush Time [sec]
0
120
0
120
0
120
0
120
Frequency [1/min]
2
60
I-time [sec]
0.15
15
Man Breath Time [sec]
2
30
2
30
O
2
[%]
21
100
21
100
21
100
21
100
O
2
Flush [%]
23
100
23
100
23
100
23
100
PDUO [mbar]
5
15
Pmanual [mbar]
5
15