130 20 Electromagnetic compatibility statement
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Guideline and Manufacturer Declaration Electromagnetic Immunity
The device fabian +nCPAP evolution is intended for operation in the environment described below. The
customer or operator of the fabian +nCPAP evolution apparatus should ensure it is operated in this type of
environment.
Immunity
Testing
Compliance Level
Electromagnetic Environment Guideline
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Air: ±2kV, ±4kV, ±6kV, ±8kV, ±15kV
Contact: ±2kV, ±4kV, ±6kV, ±8kV
Floors should be wooden or concrete or tiled with
ceramic tiles. For flooring made from synthetic
materials, the Relative Air Humidity must be no
less than 30%.
Electric Fast
Transients & Burst
IEC 610004-4
AC Power Leads: ±2kV
SIP / SOP & Control ports: ±1kV
(All cable lengths tested)
The Mains quality should correspond to typical
business or hospital environments.
Surges
IEC 61000-4-5
Differential Mode (Lead-to-Lead)
±0.5kV, ±1kV
Common Mode (Lead-to-Ground)
±0.5kV, ±1kV, ±2kV
The Mains quality should correspond to typical
business or hospital environments.
Power Frequency
Magnetic Fields
IEC 61000-4-8
30 A/m rms
Magnetic areas at the Mains frequency should
correspond to typical values in business and
hospital environments.
Voltage Dips,
Interruptions &
Voltage Variations
IEC 61000-4-11
0% UT
(100% dip of UT) for 0.5 cycles at
various phase angles; 0, 45, 90,
135, 180, 225, 270, 315
0% UT
(100% dip of UT) for 1 cycle at a
phase angle of 0.
40% UT
(60% dip of UT) for 5 cycles at a
phase angle of 0.
70% UT
(30% dip of UT) for 25 /30 cycles at
a phase angle of 0.
0% UT
(100% dip of UT) for 250 /300
cycles at a phase angle of 0.
The Mains quality should correspond to typical
business or hospital environments. If the operator
of the fabian +nCPAP evolution requires
continued use in the event of a power failure we
recommend connecting the fabian +nCPAP
evolution to an uninterruptible Mains supply or a
Battery.
REMARK : UT is the Mains alternating voltage before applying the Test Level.