4 System functions and displays 19
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
The alarms can be set between the following ranges:
Apnea [sec]
(in CPAP, NCPAP, DUOPAP, SIMV, SIMV+PSV, PSV modes)
2 to 20, OFF
CPAP [mbar]
(in CPAP mode)
Upper limit
9 to 40
Lower limit
10 to 39
CPAP [mbar]
(in NCPAP mode)
Upper limit
4 to 30
Lower limit
5 to 29
Frequency [bpm]
(in CPAP, IPPV, SIPPV, SIMV, SIMV+PSV, PSV, NCPAP,
DUOPAP modes)
Upper limit
10 to 75, OFF
Leak [%]
(in CPAP, IPPV, SIPPV, SIMV, SIMV+PSV, PSV)
Upper limit
OFF, 10 to 50
Minute Volume [liter]
(in CPAP, IPPV, SIPPV, SIMV, SIMV+PSV, PSV modes)
Upper limit
Neo. 0.01 to 7.0
Ped. 0.01 to 10.0
Lower limit
Neo. OFF, 0.01 to 6.9
Ped. OFF, 0.01 to 9.0
PEEP [mbar]
(in DUOPAP mode)
Lower limit
5 to 29
PEEP [mbar]
(in IPPV, SIPPV, SIMV, SIMV+PSV, PSV modes)
Lower limit
10 to 89
PIP [mbar]
(in DUOPAP mode)
Upper limit
4 to 30
Lower limit
OFF, 5 to 29
PIP [mbar]
(in IPPV, SIPPV, SIMV, SIMV+PSV, PSV modes)
Upper limit
1 to 90
Lower limit
OFF: 0 to 89
WARNING:
The respiratory rate and minute volume monitors are calculated as averages using an
update period of 6 seconds. In the event that the monitored breath rate falls below
11 bpm, these values are not reported.