24 4 System functions and displays
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
List of locked values:
Parameter
Neonatal
Pediatrics
CPAP [mbar]
> 10
> 10
E-Flow [L/min]
> 20
Not
blocked
Flow [L/min]
> 5
> 5
I:E [Ratio]
I : E > 1:1
I : E > 1:1
I-Flow [L/min]
> 20
Not
blocked
PEEP [mbar]
< 2
< 2
PEEP [mbar]
> 10
> 10
PInsp [mbar]
> 25
> 25
PPSV + PEEP [mbar]
> 25
> 25
V guarant [mL]
> 30
> 30
V limit [mL]
> 30
> 30

Table of Contents