58 7 Alarms
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Access the Alarm Limits menu by pressing the Alarm Limits button.
The Alarm Limits menu allows all alarm
limits to be manually or automatically
adjusted.
In the event of an alarm, a notice
appears in the Information bar and a
crossed out Bell symbol indicates the
alarm cause.
In addition, an audible alarm sounds,
which can be silenced for two minutes
by pressing the Alarm Silence button.
The Alarm limit can be adjusted
manually (1).
The Alarm Limits menu automatically
focuses on the corresponding
parameter.
The limit may also automatically be
adjusted by pressing the button
underneath Autoset (2).
Figure 7-2: Alarm limits menu
After the Alarm Limit has been adjusted or there are no more active alarms, the bell symbol
will turn Grey.
The Visual Alarm message must be reset by pressing the Alarm Silence button.