62 7 Alarms
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
Low physical
memory -
please reboot!
System failure
System error
Contact your local
ACUTRONIC Medical
Systems Distributor
I
2
Air supply
pressure
System alarms
Air supply pressure
too low.
Ensure pressure
above 2 bar.
I
2
Oxygen supply
pressure
System alarms
Oxygen supply
pressure too low.
Ensure pressure
above 2 bar.
I
2
Safety relay
defect
System failure
Safety relay defective.
Contact your local
ACUTRONIC Medical
Systems Distributor
I
2
Tube
Occlusion
Breathing
Circuit alarms
Expiration tube
blocked/kinked
Inspiration tube
blocked/kinked
Excessive
deviation between
inspiration and
expiration
pressure sensor.
NOTE: Coughing and
crying can cause a
false tube occlusion
alarm.
Check ventilation
tubes.
Check pressure
sensor connection
(proximal).
Contact your local
ACUTRONIC
Medical Systems
Distributor
I
2
Voltage
monitoring
System failure
System error
Contact your local
ACUTRONIC Medical
Systems Distributor
I
2
Battery defect
Electrical
power
No power or too less
power from battery.
Change battery,
Check fuse.
I
2
Charge
battery
(<15min)
Electrical
power
Ventilator is not
connected to
mains supply.
Remaining battery
time < 15 min.
Switch to mains
supply.
The device
continues in
battery mode
without
interruption.
Alarm can be
silenced with
Alarm Silence
button.
I
2
Depleted
battery!
Electrical
power
Remaining battery
time < 1 min.
Ventilation can no
longer continue on
battery power
Immediately switch to
Mains supply or
Access alternative
ventilation
I
2