74 10 Ventilation parameters
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Ventilation Parameters
Parameter
Mode
Description
Max. time
manual breath
• All modes
Maximum time manual breath*
Maximum duration allowed for manual breaths.
Adjustable from 2 – 30s.
*In section 16: Setting ranges and parameters
O
2
• All modes
Oxygen concentration
Inspiratory oxygen concentration setting
O
2
Flush
• All modes
Oxygen flush
Perform O
2
flush / oxygen boost
(Preoxygenation for max. 2 minutes)
P Backup
• PSV
Inspiratory pressure
Inspiratory pressure during mandatory inspiration in PSV
mode.
P DUO
• duoPAP
Upper inspiratory pressure in duoPAP mode.
The lower pressure level is selected in duoPAP mode via the
CPAP parameter.
P insp.
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
Inspiratory pressure
P manual
• CPAP
• nCPAP
Manual Inspiratory pressure
Applying a manual breath
Next breath is possible after a block period of 200 ms.
P PSV
• SIMV + PSV
• PSV
Inspiratory pressure PSV
Inspiratory pressure during inspiration in PSV mode.
PEEP
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
Positive end-expiratory pressure
Pressure at the airway connection port at the end of
expiration.
Pmax
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
Maximum Pressure
Maximum inspiratory pressure during volume guarantee.
Pressure Rise
Control
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
I-Flow or Rise Time
Will allow for use of either I-flow or Rise time based pressure
rise control during inspiration