DD+DIS155
.11E
Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.5 Diagnose & Repair
2.5.1 Error Code Explanation
Purpose
The menu “Explain Error Code” list the description for the digitizer error codes which
are reported by the error viewer.
Usage
(1) Select service menu: <Explain
error code>
(2) Enter the error code and click button:
<Get Error Description>
The error description is displayed.
Figure 15
NOTE:
The digitizer issues two types of error codes:
Error Code Type Example Example code to be entered in
the Service User Interface
Pure service message ARC1001 (25613) ARC1001*
R&D internal error code 24588 24588
*Also the internal error code (in this example "25613") can be looked up.
Edition 1, Revision 1 CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 20 of 31
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential

Table of Contents

Other manuals for AGFA CR 10-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AGFA CR 10-X and is the answer not in the manual?

AGFA CR 10-X Specifications

General IconGeneral
BrandAGFA
ModelCR 10-X
CategoryMedical Equipment
LanguageEnglish

Related product manuals