DD+DIS155
.11E
Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.7.2 Restore
Purpose
The menu “Restore” is required to restore a certain configuration of the digitizer.
Usage
(1) Select service menu: <Restore>
(2) Click: <Read from USB> or <Read from PC>
(3) Select the backup file. It is a file with following syntax:
BAC_CR10-5151_<Serial number>DDMMMYYYY_HH-MM-SS.ZIP
Figure 26
Edition 1, Revision 1 CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 29 of 31
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential