DD+DIS155
.11E
Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.5.4 Endurance Run Cycle
Purpose
The menu “Endurance Run Cycle” allows to run up to 300 cassette cycles with or
without scanning and erasure:
This can be helpful when investigating sporadic image plate handling failures.
Usage
(1) Select service menu: <Diagnose
cycle>: The Error Viewer window
opens.
(2) Click: <Confirm>
Figure 19
(3) Enter the number of cycles
(4) Select <NO> for Scanning (i.e. laser
beam switched off during scan) and /
or Erasure if required.
(5) Put a cassette into the digitizer.
(6) Click: <Start>
(7) Click <Stop> to interrupt before the
number of adjusted cycles is
completed.
Figure 20
IMPORTANT:
The Error Viewer and related confirm / cancel window only pops up at the
NX workstation. This means, that in any case, entering the Endurance Run
window needs to be confirmed at the NX workstation.
Only leave the Endurance Run window by the CANCEL button. Leaving it via
browser back button requires a reboot of the digitizer to allow further service
activities.
Edition 1, Revision 1 CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 23 of 31
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential

Table of Contents

Other manuals for AGFA CR 10-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AGFA CR 10-X and is the answer not in the manual?

AGFA CR 10-X Specifications

General IconGeneral
BrandAGFA
ModelCR 10-X
CategoryMedical Equipment
LanguageEnglish

Related product manuals