DD+DIS170
.11E
Installation Planning
123
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3.2 Guidelines for Placing the Digitizer
Observe following guidelines for the installation place of the digitizer:
ï‚· Recommended table height: 75 cm
ï‚· Due to the workflow* of the system it is recommended to install digitizer and
NX workstation close together.
* The operator has to go from digitizer to the NX workstation and back to the digitizer for each
scan cycle.
WARNING:
The device is a table-top digitizer. The structure and stability of the table used need to
be suitable in relation with the size and weight of the system. Do not use excessive
force when inserting cassettes in the digitizer as the device may slip or drop off the
table. Use a non-slip-mat below the digitizer or other anti-skid measures. The table
should not be subject to excessive shock and vibrations from other sources, as this
may disturb the operation of the digitizer.
Edition 1, Revision 1 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 11 / Page 11 of 24
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential

Table of Contents

Other manuals for AGFA CR 10-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AGFA CR 10-X and is the answer not in the manual?

AGFA CR 10-X Specifications

General IconGeneral
BrandAGFA
ModelCR 10-X
CategoryMedical Equipment
LanguageEnglish

Related product manuals