DD+DIS155
.11E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
5.4 Verifying the Installation
NOTE:
Verifying the installation means: Expose a flatfield and evaluate the image quality.
Following items are verified with the check:
ï‚· The configuration was successful: Images arrive at the workstation.
ï‚· Image quality is OK.
(1) Erase a cassette.
install_poster_12.cdr
B
C
D
A
Insert cassette with
black side facing up.
Press the "Erase" button.
Wait approx. 3 minutes.
Remove cassette.
Figure 61
(2) Use the erased cassette to
expose a flatfield.
ï‚· The black cassette side
must be facing up.
ï‚· The collimated area
should be larger than the
cassette.
Black side
facing up
Collimated
Area
1. Insert CU Filter
2. Expose Cassette:
Large Focus
12 mAs
75 kVp
3. Rotate cassette by 180°
4. Expose again
5. Remove CU Filter
1,3 m
install_poster_013.cdr
Figure 62
Edition 1, Revision 4 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 1 / Page 43 of 47
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential