Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.6 / Page 4 of 10
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential
1 Overview of Adjustments and Calibrations
The following tables give an overview of adjustments and calibrations for the digitizer.
Adjustment Type Use Case Section
None Not applicable: After spare parts exchange no
adjustment is required.
--
Calibration Type Use Case Section
Shading calibration
ď‚· After replacement of the PMT with Light
Collector
ď‚· After replacement of the Optic Module
ď‚· Whenever stripes in image plate run direction
are visible which cannot be removed by
cleaning with the cleaning brush