DD+DIS155
.11E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
4.6 Checking Software Version
Item
Type Number:
Subtype Number:
Serial Number:
Manufacture Date:
Installation Date:
Software Version:
Device Info
Value
5151
100
1001
02.01.2012
15.01.2012
ARC_1013
E
XA
M
PL
E
(1) Select service menu
Device Info and check
the digitizer software
version.
(2) Perform a software update
if a newer software is
available. Refer to the
instructions enclosed to the
software ZIP file.
Figure 15
NOTE:
The latest software is available from the Agfa HealthCare Library:
Computed Radiography => CR Digitizer => CR 10-X / CR 12-X => Software
4.7 Performing further Activities depending on local Regulations
Enclosed a list of possible activities depending on local regulations:
This list is just an example.
# Activity Reference
1 Electrical check IEC 62353:2007 Test Documentation on
the Agfa HealthCare Library (Extranet Link
/
Intranet Link
)
2 Acceptance Test Example for Germany: DIN6868
(1) Contact your local service manager concerning the activities which are required
due to local regulations.
4.8 Customer Training
(1) Hand-over the user manuals.
(2) Train the customer in following activities:
ï‚· Operating the digitizer, see section 4.8.1.
ï‚·
Usage of plates and cassettes, see section 4.8.2.
Edition 1, Revision 4 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 1 / Page 18 of 47
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential