EasyManuals Logo

Agilent Technologies 1100 Series User Manual

Agilent Technologies 1100 Series
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
1100 Series DAD and MWD User Manual 65
Using the Detector 4
6 Allow the detector to warm up of at least 60 minutes to provide a stable
baseline (see example in Figure 34 and Table 13).
.
NOTE
For reproducible chromatography, the detector and lamp should be on for at least one hour.
Otherwise the detector baseline may still drift (depending on the environment). See also
section “Wander / Drift Problems Due to Temperature Changes” in the Service Manual.
Figure 34 Stabilization of Baseline (both lamps turned on at the same time)
Table 13 Baseline drift after lamp turn on (example from Figure 34)
Time [minutes] Drift [mAU/hr]
17 - 20 2.6
27 - 30 0.8
37 - 40 0.4
47 - 50 0.2
57 - 60 < 0.2

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals