1260 Infinity Binary Pump User Manual 49
Installing the Pump
3
Flow Connections with Solvent Selection Valve
2 If available, place the online degasser on top of the pump.
3 Place the solvent cabinet on top of the module.
4 Set the four bottles into the solvent cabinet and screw a bottle head
assembly onto each bottle.
5 Connect the solvent tubes from the bottle head assemblies to the inlet
connectors A1, A2, B1 and B2 of the solvent selection valve. Make sure
to use the brown bottle for the aqueous solvent (usually channel A1).
6 Label the tubes accordingly using the supplied stickers and fix the
tubes in the clips of solvent cabinet and binary pump.
7 Hold the waste tubing with a piece of sandpaper and push it onto the
purge valve outlet. Place the end into your waste system.
8 If the pump is not part of an Agilent 1260 Infinity system stack or
placed on the bottom of a stack, connect the waste tube to the waste
outlet of the pump leak handling system.
9 Connect the pump outlet capillary (pump to injection device) to the
outlet of the purge valve.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals