52 1260 Infinity Binary Pump User Manual
3 Installing the Pump
Flow Connections without Solvent Selection Valve
2 Place the solvent cabinet on top of the module.
3 Place the bottles into the solvent cabinet and place a bottle head
assembly into each bottle.
4 Connect the solvent tubes from the bottle head assemblies to the inlet
adapters of the active inlet valves. Fix the tubes in the clips of solvent
cabinet and binary pump.
5 Hold the waste tubing with a piece of sandpaper and push it onto the
purge valve outlet. Place the end into your waste system.
6 If the pump is not part of an Agilent 1260 Infinity system stack or
placed on the bottom of a stack, connect the waste tube to the waste
outlet of the pump leak handling system.
7 Connect the pump outlet capillary (pump to injection device) to the
outlet of the purge valve.