68 1260 Infinity Binary Pump User Manual
4 Using the Pump
Setting up the Agilent 1260 Infinity Binary Pump G1312B with the Instrument Control Interface
Tabl e 6 Bottle Filling Parameters
Parameter Limits Description
• Total 
Volume
0 – 1000 L
default: 0 L
This is the capacity (maximum possible volume) in liter of the solvent bottle. In 
combination with the actual volume, this parameter is used for calculating and 
displaying the relative liquid level.
•Actual 
Volume
0 – 1000 L
default: 0 L
After filling the solvent bottles, enter the actual volumes into these boxes. The 
Actual Volume must not be larger than the Total Vo lume of the bottle.
• Prevent 
analysis.......
default: unchecked If this option is checked, the pump won’t start a new run if the solvent level in 
one or more bottles is below the minimum volume. Enter a minimum volume in 
liter, which considers the position of the solvent inlet and size/shape of the 
solvent bottle such that no air is drawn if the actual volume gets close to this 
limit.
• Turn pump 
off...
default: unchecked If this option is checked, the pump will turn off before air is aspirated. However, 
the residual solvent volume has been calculated for 1 L solvent bottles and may 
be too small for large bottles or other vessels.