68 1260 Infinity Binary Pump User Manual
4 Using the Pump
Setting up the Agilent 1260 Infinity Binary Pump G1312B with the Instrument Control Interface
Tabl e 6 Bottle Filling Parameters
Parameter Limits Description
Total
Volume
0 – 1000 L
default: 0 L
This is the capacity (maximum possible volume) in liter of the solvent bottle. In
combination with the actual volume, this parameter is used for calculating and
displaying the relative liquid level.
•Actual
Volume
0 – 1000 L
default: 0 L
After filling the solvent bottles, enter the actual volumes into these boxes. The
Actual Volume must not be larger than the Total Vo lume of the bottle.
Prevent
analysis.......
default: unchecked If this option is checked, the pump won’t start a new run if the solvent level in
one or more bottles is below the minimum volume. Enter a minimum volume in
liter, which considers the position of the solvent inlet and size/shape of the
solvent bottle such that no air is drawn if the actual volume gets close to this
limit.
Turn pump
off...
default: unchecked If this option is checked, the pump will turn off before air is aspirated. However,
the residual solvent volume has been calculated for 1 L solvent bottles and may
be too small for large bottles or other vessels.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals