EasyManua.ls Logo

Agilent Technologies 1260 Infinity - Page 89

Agilent Technologies 1260 Infinity
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1260 Infinity Binary Pump User Manual 89
Optimizing Performance
5
How to Optimize the Compressibility Compensation Setting
Tabl e 8 Solvent Compressibility
Solvent (pure)
Compressibility (10
-6
/bar)
Acetone 126
Acetonitrile 115
Benzene 95
Carbon tetrachloride 110
Chloroform 100
Cyclohexane 118
Ethanol 114
Ethyl acetate 104
Heptane 120
Hexane 150
Isobutanol 100
Isopropanol 100
Methanol 120
1-Propanol 100
To l u e n e 8 7
Water 46