1290 Infinity II High-Speed Pumps User Manual 103
6 Troubleshooting and Diagnostics
Pump Leak Rate Test
Outlet valve not properly assembled
Retighten the outlet valve
Check the position of the gold seal
Leaky piston seals and/or position-dependent leaks on the piston
Remove the SW tubes from the support ring and check for leaks
Replace the piston seals and clean the pistons
Ensure that seals are lubricated when pushed in
Use abrasive mesh >5000 grit