1290 Infinity II High-Speed Pumps User Manual 204
8Maintenance
Replace Parts of the High Pressure Filter Assembly (G7120A)
3 Replace the filter frit and seal cap as desired. Please
note the correct orientation of the filter frit.
4 Re-install the filter assembly using the torque wrench
(14 mm) set to approx. 16 Nm.
5 Reconnect the capillary from the purge valve to the high
pressure filter assembly.
Filter frit
Seal cap
16 Nm
approx.