1290 Infinity II High-Speed Pumps User Manual 206
8Maintenance
Replace Parts of the High Pressure Filter Assembly (G7132A)
3 Replace the filter frit and seal cap as desired. Please
note the correct orientation of the filter frit.
4 Reinstall the filter assembly using the torque wrench
(14 mm hex bit) set to approx. 16 Nm.
5 Reconnect the capillary from the purge valve to the
high-pressure filter assembly.
Filter frit
Seal cap
16 Nm