1290 Infinity II High-Speed Pumps User Manual 31
2 Site Requirements and Specifications
Performance Specifications
Solvent selection valve Internal 4-solvent selection
valve included. External 2 x 12
solvent valve as option, fully
integrated in the pump control
interface.
Integrated degassing unit Number of channels: 2
Internal volume per channel:
1.5 mL
Materials in contact with sol-
vent
TFE/PDD copolymer, FEP,
PEEK, PPS, MP35N
Automatic purge valve Included
Active Seal Wash Included
Intelligent System Emulation
Technology (ISET)
Included
Instrument control LC & CE Drivers Rev. 3.3 or
above
Instrument Control Framework
(ICF) A.02.03 or above in Emula-
tion Mode only
InfinityLab Companion
(G7108AA) with firmware
D.07.33 or above
Lab Advisor software B.02.16 or
above
For details about supported
software versions refer to the
compatibility matrix of your ver-
sion of the LC & CE Drivers
Communication Controller Area Network (CAN),
RS232C, APG remote: ready,
start, stop, and shutdown sig-
nals, LAN
Safety features and mainte-
nance
Leak detection, safe leak han-
dling, leak output signal for
shutdown of the pumping sys-
tem.
Low voltage in major mainte-
nance areas.
Extensive diagnostics, error
detection and display with Agi-
lent Lab Advisor software.
Table 4 Performance specifications G7132A
Type Specification Comments