Amplitude Accuracy (continued)
Measurement Lower Limit Upper Limit
(dBfs)
Measured Value
(dBfs)
Pass/Fail
35 dBVrms, Ch 2 35.15 34.85
35 dBVrms, Ch 3 35.15 34.85
35 dBVrms, Ch 4 35.15 34.85
27 dBVrms, Ch 1 27.15 26.85
27 dBVrms, Ch 2 27.15 26.85
27 dBVrms, Ch 3 27.15 26.85
27 dBVrms, Ch 4 27.15 26.85
11 dBVrms, Ch 1 11.15 10.85
11 dBVrms, Ch 2 11.15 10.85
11 dBVrms, Ch 3 11.15 10.85
11 dBVrms, Ch 4 11.15 10.85
1 dBVrms, Ch 1
0.85 1.15
1 dBVrms, Ch 2
0.85 1.15
1 dBVrms, Ch 3
0.85 1.15
1 dBVrms, Ch 4
0.85 1.15
9 dBVrms, Ch 1
8.85 9.15
9 dBVrms, Ch 2
8.85 9.15
9 dBVrms, Ch 3
8.85 9.15
9 dBVrms, Ch 4
8.85 9.15
19 dBVrms, Ch 1
18.85 19.15
19 dBVrms, Ch 2
18.85 19.15
19 dBVrms, Ch 3
18.85 19.15
19 dBVrms, Ch 4
18.85 19.15
27 dBVrms, Ch 1
26.85 27.15
27 dBVrms, Ch 2
26.85 27.15
27 dBVrms, Ch 3
26.85 27.15
27 dBVrms, Ch 4
26.85 27.15
Serial Number:_______________________Report Number:____________________