Press the following keys:
[
Rtn ]
[
CHANNEL 3 ADJUSTMNT ]
[
25 kHz FLATNESS ]
Wait for the analyzer to set up the adjustment. The analyzer is ready when
the adjustment message appears on the screen.
While monitoring the Yr: value, adjust A2 R615 in the lower slot for a
marker value of 0 0.1 dB.
Press the following keys:
[
Rtn ]
[
CHANNEL 4 ADJUSTMNT ]
[
25 kHz FLATNESS ]
Wait for the analyzer to set up the adjustment. The analyzer is ready when
the adjustment message appears on the screen.
While monitoring the Yr: value, adjust A2 R615 in the upper slot for a
marker value of 0 0.1 dB.
Adjusting the Analyzer Agilent 35670A
To adjust filter flatness
5-20