Loading...
Actual
Reading
R14 Adjusted
Voltage
Actual
Reading
R14 Adjusted
Voltage
Actual
Reading
R14 Adjusted
Voltage
V (TP11) V (TP15) V (TP11) V (TP15) V (TP11) V (TP15)
0.201 0.281 0.236 0.330 0.271 0.379
0.202 0.283 0.237 0.332 0.272 0.381
0.203 0.284 0.238 0.333 0.273 0.382
0.204 0.286 0.239 0.335 0.274 0.384
0.205 0.287 0.240 0.336 0.275 0.385
0.206 0.288 0.241 0.337 0.276 0.386
0.207 0.290 0.242 0.339 0.277 0.388
0.208 0.291 0.243 0.340 0.278 0.389
0.209 0.293 0.244 0.342 0.279 0.391
0.210 0.294 0.245 0.343 0.280 0.392
0.211 0.295 0.246 0.344 0.281 0.393
0.212 0.297 0.247 0.346 0.282 0.395
0.213 0.298 0.248 0.347 0.283 0.396
0.214 0.300 0.249 0.349 0.284 0.398
0.215 0.301 0.250 0.350 0.285 0.399
0.216 0.302 0.251 0.351 0.286 0.400
0.217 0.304 0.252 0.353 0.287 0.402
0.218 0.305 0.253 0.354 0.288 0.403
0.219 0.307 0.254 0.356 0.289 0.405
0.220 0.308 0.255 0.357 0.290 0.406
0.221 0.309 0.256 0.358 0.291 0.407
0.222 0.311 0.257 0.360 0.292 0.409
0.223 0.312 0.258 0.361 0.293 0.410
0.224 0.314 0.259 0.363 0.294 0.412
0.225 0.315 0.260 0.364 0.295 0.413
0.226 0.316 0.261 0.365 0.296 0.414
0.227 0.318 0.262 0.367 0.297 0.416
0.228 0.319 0.263 0.368 0.298 0.417
0.229 0.321 0.264 0.370 0.299 0.419
0.230 0.322 0.265 0.371 0.300 0.420
0.231 0.323 0.266 0.372 0.301 0.421
0.232 0.325 0.267 0.374 0.302 0.423
0.233 0.326 0.268 0.375 0.303 0.424
0.234 0.328 0.269 0.377 0.304 0.426
0.235 0.329 0.270 0.378 0.305 0.427
Adjustments 5
Table 11 Setpoint voltages at A3TP11 and A3TP15 – short body tube*
* The setpoint voltages at A3TP11 and A3TP15 listed in this table apply to the 05517-60255 or -60256 short body tube with the 05517-60031
A3 Board, and the 05517-60253 or -60254 short body tube with the 05517-60032 A3 Board.