Verifying Performance
µECD Checkout
Released: March 2004 6850 Series II GC User Information page 70 of 256
Offset checkout
Display the signal output to determine background offset. A stable offset value
less than 1000 display units (inclusive) is acceptable.
If the offset is < 0.5 display units, verify that the electrometer is on. If the
offset is still < 0.5 display units, your detector requires service.
If offset is > 1000 display units, there may be chemical contamination
contributing to the signal. Perform a thermal cleanout (see “Thermal
cleaning” on page 207 for details). If bakeout does not give an acceptable
signal, check gas purity. Use higher purity gases and/or install traps.
Rate 2 100°C/min
Final temp 2 250°C
Final time 2 0 min
Purge time 0.75 min
Purge flow 60 mL/min
Detector
Temperature 300°C
Anode purge flow (N
2
) 60 mL/min
Makeup flow (N
2
)25±2mL/min
Offset Should be < 1000 display counts
Oven
Initial temp 80°C
Initial time 0 min
Rate 1 15°C/min
Final temp 180°C
Final time 10 min
Table 19. µECD Checkout Conditions (Method ECDCKO)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals