Verifying Performance
µECD Checkout
Released: March 2004 6850 Series II GC User Information page 69 of 256
ChemStation setup
1. Bring up the proper on-line instrument for the 6850 to be tested. Make
these settings:
2. Enter the parameter values listed in Table 19.
3. Define the column.
4. Save the method. We suggest using the name ECDCKO.
Path Setting
View/Run Method/Control View/Instrument/
Edit Parameters/Options
Keep instrument keyboard locked after method is
loaded? = No
View/Run Method/Control View/Instrument/
Edit Parameters/Signals
Assign Signal 1 to the detector
Choose Save Data All
Data Rate = 20 Hz
Table 19. µECD Checkout Conditions (Method ECDCKO)
Column and sample
Type HP-1, 30 m × 0.32 mm × 0.25 µm part number 19091Z–413E
Sample µECD checkout 18713-60040
Injection volume 1 µL
Inlets
Temperature 200°C Purged packed
250°C Split/splitless
Oven Track (cool on-column)
80°C PTV (see below)
Inlet pressure 25 psi (constant pressure, helium)
Split/splitless
Mode Splitless
Purge flow 60 mL/min
Purge time 0.75 min
PTV
Mode Splitless
Inlet temperature 80°C
Initial time 0.1 min
Rate 1 720°C/min
Final temp 1 350°C
Final time 1 2 min

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals