4 Operating the 850-DS
850-DS Operator’s Manual 41
Full Media Change Signifies that you will be removing the dosage and performing a complete
replacement of media. A dialog box will display to prompt you to begin and to
keep track of the total elapsed time. This option will pause the elapsed time of
the dissolution method during the change.
Media Addition Signifies that you will be adding a quantity of media to each vessel. A dialog will
display to prompt you to begin as well as to keep track of the total elapsed time.
This option will continue the elapsed time of the dissolution method during the
change.
Enable Cleaning
Cycle
When enabled, executes a cleaning cycle after the test has finished. The
parameters for the cleaning routine are taken from the stored values on the
Cleaning screen (“Cleaning” on page 61). A validated cleaning method
should be performed after each dissolution test. This enables the 850-DS
to perform at a high level for an extended period of time while reducing
ongoing maintenance costs.
Clear Erases all values entered and returns them to default values.
Option Function