7 Maintaining the COC Inlet
To Change a Septum on the COC Inlet
110 Maintaining Your GC
To Change a Septum on the COC Inlet
1 Gather the following:
Replacement septum, see “Consumables and Parts for the Cool On-Column Inlet” on
page 102.
Tweezers
A thin wire (0.2-inch diameter) for removing septum from inlet
Lint-free gloves
2 Launch the GC maintenance wizard: Maintenance > Inlets, select the inlet, then select
Perform Maintenance > Replace Septum > Start Maintenance. The wizard will walk
through the steps needed to replace the septum (those same steps are repeated below).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3 Replace the septum.
If you are using a septum nut, grasp the knurling and unscrew. Remove the old septum
with tweezers.
Use tweezers to install a new septum. Push the septum into the septum nut until
properly seated. Firmly tighten the nut.
If you are using a cooling tower, grasp the three rings and unscrew. The spring and
duckbill septum may pop out of the inlet when you remove the cooling tower. Be careful
not to lose them. If they do not pop out, use a thin wire to remove them from the inlet.
Insert the replacement duckbill septum into the spring and place in the inlet. Reattach
the cooling tower assembly, then finger-tighten.
For 250/320-µm
automated injections
For 530-µm automated
injections

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals