EasyManua.ls Logo

Agilent Technologies 8890 - To Install a Capillary Column in the ECD

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14 Maintaining the ECD
To Install a Capillary Column in the ECD
236 Maintaining Your GC
To Install a Capillary Column in the ECD
1 Gather the following:
Ferrule, see “Consumables and Parts for the ECD” on page 230.
Column nut
Septum
Column
1/4-inch, 5/16-inch, and 9/16-inch wrenches
Column cutter
Lint-free gloves
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3 Load the inlet maintenance method and wait for the GC to become ready.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4 Place a septum (if the column id is 0.2 mm), capillary column nut, and ferrule on the
column.
5 Score the column using a glass scribing tool. The score must be square to ensure a clean
break.
Ferrule
Column nut
Septum

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals