11 Maintaining the FID
To Install a Capillary Column in the FID
184 Maintaining Your GC
To Install a Capillary Column in the FID
1 Gather the following materials, see “Consumables and Parts for the FID” on page 180:
Column
Ferrule(s)
Column nut
Column cutter
1/4-inch open-end wrench
Septum
Isopropanol
Lab tissue
Lint-free gloves
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3 Place a septum (if the column id is 0.1 mm), capillary column nut, and ferrule on the
column.
4 Score the column using a glass scribing tool. The score must be square to ensure a clean
break.
Ferrule
Column nut
Septum

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals