2Removing Covers
To Remove the Detector Top Cover
18 Maintaining Your GC
To Remove the Detector Top Cover
This cover protects the detectors, valve box, and valve assembly. To remove the detector top
cover:
1 Turn off the GC and unplug the power cord.
2 Remove the detector cover by raising the cover up, then firmly lift up on the right side of the
cover to free it from the GC. Set the cover aside.
3 When finished with the maintenance procedure, place the detector cover vertically on the
GC and insert the left side first, then gently snap in the right side onto the GC.
WARNING
Be careful! With the power switch on, potentially dangerous voltages also exist:
All electronics boards in the instrument.
•The internal wires and cables connected to these boards.
The wires for the oven heater.
The covers shield these parts as the parts may be hot enough to cause burns. Never remove
a cover unless the instrument is unplugged.
CAUTION
Do not force the cover, either when installing it or closing it. This could break the plastic
parts.
Figure 2. 8890 GC covers
Detector coverPneumatics cover
Injection port cover
Left side cover
Power switch
Pneumatics cover
release button (on
both sides)

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals