15 Maintaining the FPD+
To Remove the FPD+ Cover
250 Maintaining Your GC
To Remove the FPD+ Cover
1 Gather the following:
T-20 Torx screwdriver
2 Turn off the flame, then turn off the GC.
CAUTION
When turning off the GC, first turn off the flame to prevent condensation from dripping into
the jet and column.
WARNING
Be careful! The detector may be hot enough to cause burns. If the detector is hot, wear
heat-resistant gloves to protect your hands.
3 Open the FPD detector top cover.
4 Loosen the screws securing the FPD cover to the top of the detector.
5 For a single wavelength FPD, remove the two screws at the bottom left of the cover.
6 Lift the cover off the detector.