EasyManua.ls Logo

Agilent Technologies 8890 - To Check the FID Baseline

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 Maintaining the FID
To Check the FID Baseline
198 Maintaining Your GC
To Check the FID Baseline
1 With the column installed, load your checkout method.
2 Set the oven temperature to 35 °C.
3 Show the detector output signal in Status. On the GC touchscreen go to Home > Status
listing > +Add. Select Detector’s Output from the drop-down list, then touch Add.
4 When the flame is lit and the GC is ready, verify that the output is stable and < 20 pA.
If the output is not stable or > 20 pA, the system or gas may be contaminated. If this
contamination is isolated to the detector, then bakeout the FID. See “To Bakeout the
FID” on page 200.

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals