7 Maintaining the COC Inlet
To Install an Insert on the COC Inlet
112 Maintaining Your GC
To Install an Insert on the COC Inlet
1 Gather the following:
Lint-free gloves
Replacement insert, see “Consumables and Parts for the Cool On-Column Inlet” on
page 102.
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3 Remove the column from the inlet.
4 Locate the septum nut or cooling tower assembly at the top of the inlet and remove it. If
the septum remains in the septum nut, do not remove it unless you want to change it. If
necessary, replace the existing septum or duckbill with a new one. See “To Change a
Septum on the COC Inlet” on page 110.
5 Remove the spring from the inlet with an extraction wire (or tweezers) and set it aside. Be
careful not to lose or damage it because you will use the spring to keep the new insert in
position.
6 Remove the existing insert from the inlet by gently pushing it out from below with a wire or
piece of column. Store the insert for possible later use.

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals