EasyManuals Logo

Agilent Technologies 8890 Maintaining Manual

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
13 Maintaining the NPD
To Install a Capillary Column in the NPD
Maintaining Your GC 213
To Install a Capillary Column in the NPD
1 Gather the following materials:
Column
Ferrule(s), see “Consumables and Parts for the NPD” on page 210.
Column nut
Column cutter
1/4-inch open-end wrench
Septum
Isopropanol
Lab tissue
Lint-free gloves
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3 Place a septum (if the column id is 0.1 mm), capillary column nut, and ferrule on the
column.
4 Score the column using a glass scribing tool. The score must be square to ensure a clean
break.
Ferrule
Column nut
Septum

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals