EasyManua.ls Logo

Agilent Technologies 8890 - Page 214

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13 Maintaining the NPD
To Install a Capillary Column in the NPD
214 Maintaining Your GC
5 Break off the column end by supporting it against the column cutter opposite the scribe.
Inspect the end with a magnifying loupe to make certain there are no burrs or jagged
edges.
6 Wipe the column walls with a tissue dampened with isopropanol to remove fingerprints
and dust.
7 Install the capillary column.
If the column id is greater than 0.1 mm:
a Gently insert the column into the detector until it bottoms; do not attempt to force it
further.
b Finger-tighten the column nut, then withdraw the column about 2 mm. Tighten the nut
an additional 1/4 turn with a wrench.
Good
Bad

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals