EasyManua.ls Logo

Agilent Technologies 8890 - To Replace the 7693 A Injector Needle Support Guide

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Maintaining the COC Inlet
To Replace the 7693A Injector Needle Support Guide
116 Maintaining Your GC
To Replace the 7693A Injector Needle Support
Guide
Before using a 7693A ALS to make injections onto the COC inlet, install the on-column needle
support guide.
Refer to the 7693A ALS documentation for complete details.
1 Open the injector door.
2 Remove the syringe.
3 Slide the syringe carriage up to the top position.
4 Completely remove the T-10 Torx screw from the support foot. Be careful to not let the
screw fall into the turret assembly.
5 Slide off the support foot.
6 Slide on the new support foot.
7 Replace the T-10 Torx screw and tighten.
8 Install the appropriate syringe.
9 Close the injector door.
10 Align the injector.

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals