7 Maintaining the COC Inlet
To Clean the COC Inlet
Maintaining Your GC 115
8 Remove the spring and insert from the bath and rinse thoroughly with water and methanol.
9 Dry the spring and insert with compressed air or nitrogen.
10 Install the insert. See “To Install an Insert on the COC Inlet” on page 112.
11 Install the column. See “To Install a Capillary Column with the Cool On-Column Inlet” on
page 106.
12 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.