4 Maintaining Capillary Columns
To Reverse a Column and Bakeout Contaminants
Maintaining Your GC 37
To Reverse a Column and Bakeout Contaminants
1 Gather the following:
1/4-inch wrench
Column cutter
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance and wait for the GC to become ready.
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3 Disconnect the column from the inlet and detector.
4 If necessary, cut a loop from the column. See “To Cut a Loop from a Column” on page 36.
Do not attach the column to the inlet.
5 Remove the column from the hanger and reverse its position (inlet and detector ends) and
place the column back on the hanger.
6 Attach the column to the inlet.
“To Install a Capillary Column with the Split/Splitless Inlet”
“To Install a Capillary Column with the Multimode Inlet”
“To Install a Capillary Column with the Purged Packed Inlet”
“To Install a Capillary Column with the Cool On-Column Inlet”
“To Install a Capillary Column with the Programmable Temperature Vaporization
Inlet”
“To Install a Capillary Column with the Volatiles Interface”
7 Attach your column to the detector.
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the ECD
To Install a Capillary Column to the FPD+
8 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
9 Set the column flow to the normal operating value, or set the capillary column gas velocity
to 30 cm/s.
For Multimode, Split/Splitless, PTV, and VI inlets, select split mode and set the split vent
flow to 200 mL/min.
10 Purge the column with carrier flow for at least 10 minutes before heating the oven.

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals