EasyManuals Logo

Agilent Technologies 8890 Maintaining Manual

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #237 background imageLoading...
Page #237 background image
14 Maintaining the ECD
To Install a Capillary Column in the ECD
Maintaining Your GC 237
6 Break off the column end by supporting it against the column cutter opposite the scribe.
Inspect the end with a magnifying loupe to make certain there are no burrs or jagged
edges.
7 Wipe the column walls with a tissue dampened with isopropanol to remove fingerprints
and dust.
8 Install the column.
If the column id is 200 um or more, push the column into the adapter until it stops at the
indentation. Pull it back 1 to 2 mm and tighten the column nut with one 5/16-inch wrench
on the adapter and another 1/4-inch wrench on the column nut.
If the id is less the 200 um, mark the column with a septum 70 ± 1 mm from the end.
Insert column and nut into the adapter with the septum at the rear of the column nut, and
tighten the column nut with one 5/16-inch wrench on the adapter and another 1/4-inch
wrench on the column nut.
9 After heating the detector, retighten the 9/16-inch makeup adapter nut and 1/4-inch
column nut.
10 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
Good
Bad
70 ± 1 mm
Septum

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals