8 Maintaining the MMI
To Change the Septum on the MMI
Maintaining Your GC 131
7 On the GC touchscreen, select Maintenance > Inlets > Septum injections, then touch Reset
Counter.
8 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
9 Select Finished, then select OK to exit the GC maintenance wizard.
10 Restore the analytical method.